Single-Beam NDIR Gas Analyzer with sample-switching system (drift-less type) - ZPB

Simultaneous and continuous measurement of up to 5 components among NO, SO2, CO, CO2, and O2 in incinerators or thermal power plant boilers

request quote

The product

Sample switching system offers higher stability

  • Long-term superior stability: zero drift ±0.5% FS per week
  • Compact and lightweight: 19-inch rack mount
  • Built-in magnetic O2 sensor or optional galvanic O2 sensor
  • Optional functions include: atmospheric pressure compensation, RS-485 communication, automatic calibration, alarm output, O2 correction, and more.
  • Low maintenance
request quote
tables

Specifications

PrincipleNon-dispersive infrared sensor (single beam system)
O2 sensor: Built-in magnetic or galvanic sensor or external zirconia sensor (ZFK7)
Components and measurable rangesNO: 0–50 ppm…5000 ppm
SO2: 0–50 ppm…5000 ppm
CO2: 0–50 ppm…25 vol%
CO: 0–50 ppm…5000 ppm
Built-in galvanic O2 sensor: 0–10 vol%…25 vol%
Built-in magnetic O2 sensor: 0–5 vol%…100 vol%
External Zirconia O2 Sensor: 0–5 vol%…25 vol%
No. of RangesUp to 2 ranges per component
Repeatability≤ ±0.5% FS
Linearity≤ ±1% FS
Zero Drift≤ ±0.5% FS per week
Amplitude Drift≤ ±2% FS per week
Response Time (to 90%)≤ 30 s (Loss time may vary with switching times) gas)
Analog output signal4–20 mA DC or 0–1 V DC
DisplayBacklit LCD (Japanese, English, or Chinese)
Instantaneous value, corrected instantaneous O2 value, corrected average O2 value, average O2 
Range switchingBy key operation, automatic, or by external contact input (optional)
Contact input (optional)Voltage contact input (12–24 V DC, up to 15 mA)
Remote range switching, remote start of auto calibration, remote hold, reset average
Contact output (optional)SPDT relay contact (24 V DC/1 A)
Device error, calibration error, range identification, auto calibration status, solenoid valve actuation for auto calibration, limit alarm, CO peak alarm
Atmospheric pressure correction (optional)Integrated option available
Standard functionsOutput hold, auto/manual range switching
Optional functionsAuto calibration, remote auto calibration start, remote output hold, range identification contact output, high/low limit alarm, O2 correction, corrected average O2 values, average reset contact input, peak CO alarm contact output CO
Communication (option)RS-485 (Modbus) (9-pin D-sub output): half-duplex, start-stop synchronization system
FlowmeterSupplied
Gas inlet/outletRC 1/4 or NPT 1/4 internal thread
Purge gas flow rate1 L/min (as required)
Reference gasRequired (dry N2 or dry air)
Operating environment-5°C to +45°C, RH90% or less (non-condensing)
Voltage Power Supply100–240 V AC, 50/60 Hz
Power ConsumptionApprox. 120 VA
Dimensions133 (H) × 483 (W) × 382 (D) mm
WeightApprox. 13 kg
ComplianceCE Marking

Specifications

Gas conditions

Flow rate1.0 L/min ±0.2 L/min
Temperature0–50°C
Pressure≤ 10 kPa
Dust≤ 100 μg/Nm3 with particle size ≤ 0.3 μm
MistInadmissible
HumidityBelow saturation vapor density at 2°C (sample gas and reference gas)
Corrosive gasHCl: ≤ 1 ppm