Alutal Measure & Trust

ColorPlus 3 SAK 254

ColorPlus 3 SAK 254 is a bypass absorption meter SAK 254 for online measurement of UV absorption with optional colour measurement.

request quote

The product

UV absorption is an indicator of water pollution by certain dissolved organic substances, such as aromatic compounds or humic substances from the soil. Sigrist ColorPlus 3 offers you online measurement of UV absorption with optional colour measurement. In the water industry, the ColorPlus 3 SAK 254 serves drinking water treatment.

How the ColorPlus 3 SAK 254 works

As a light source, a long-lived xenon flash lamp is used, with a very low loss of intensity over a long time. The medium is measured with two optical path lengths, from which any soiling is determined. The value measured is automatically corrected and the instrument informs the user when the measuring cell has to be cleaned.

The PVC cover can be removed without tools. Free access to the optics allows simple cleaning of the measuring cell. The instrument indicates any maintenance due.

The control unit SICON selectively displays values, graphs, status and alarms on state-of-the-art touchscreen technology and colour display. Internal data storage allows the visualization of the measurements over the last 32 days.

Applications

The ColorPlus 3 SAK 254 is used in:

  • The treatment of drinking water
  • The treatment of process water in various kinds of industries

Key features

  • On-line measurement of DOC (UV absorption) with colour (Hazen) option
  • Compensation of window fouling
  • Flow cell easy to clean without tools

Alutal, Sigrist’s exclusive distributor in Brazil for the Water and Industrial Processes segments, would be pleased to show you the possible applications on site or assist you with your projects, providing competent advice and local support in the practical use and service of the ColorPlus 3 SAK 254.

request quote
folders
tables

Technical Specifications

Parameter Value
Nominal range 0 .. 3 E
Measuring ranges 8 x freely configurable
Protection class IP67
Enclosure Stainless steel 1.4301
Measuring principle Absorption
Weight 3.4 kg
Dimensions 295 x 178 x 140 mm (W x H x D)
Wavelengths UV 1 .. 3 x 200-800 nm
Resolution 0.001 E
Units E, E/m, Hazen, GOST, mg/l
Light source Xenon flash lamp
Ambient temperature -20 °C .. +50 °C
Measuring span SAK254 0 .. 30 E/m (100 mm)
0 .. 60 E/m (50 mm)
Measuring span Hazen 0 .. 150 Hazen @390 nm (100 mm)
0 .. 300 Hazen @390 nm (50 mm)
Control Units SICON
SICON-M

100%

Energy self-sustainable

+4500

Customers around the world

ISO 9001

Certified since 2006

+5000

Parts delivered monthly