Alutal Measure & Trust

TurBiScat PM 40

TurBiScat PM 40 is an in-line turbidimeter that measures turbidity in process liquids and, optionally, colour.

request quote

The product

The TurBiScat PM 40 from Sigrist measures turbidity in process liquids and, optionally, colour. The TurBiScat thus additionally becomes an inline colour meter. In industrial processes, the TurBiScat PM 40 serves the chemical and pharmaceutical sector. In the water industry, it serves drinking water treatment.

How the TurBiScat PM 40 works

The TurBiScat PM 40 determines turbidity with optimum colour compensation using scattered light measurement. The photometer measures straylight in angles in both 90° and 25°. The operation of the sensor is monitored permanently. Calibration is easily done using a secondary reference standard.

The TurBiScat PM 40 appears in a new, fresh design with integrated coloured touchscreen display and can be identified easily in the production line. Measurement values and curves as well as data status are shown directly on point of measurement. Checking unit is integrated as well, no additional device is required. This simplifies the installation and reduces the power consumption.

Data security is a more and more important topic and has been enhanced.

Applications

The combination of stainless steel and sapphire in a sealless design allows use in virtually all process applications. With its dual angle measurement, the instrument serves:

  • Chemical and pharmaceutical processes
  • Drinking water treatment

Key features

  • High accuracy, dual angle measurement (90° + 25°), wide measurement span and optional colour measurement
  • Values and measurement curves indicated directly on point of measurement
  • Proven measurement technology, measurement method identical to laboratory instrument LabScat 2

Alutal, Sigrist’s exclusive distributor in Brazil for the Water and Industrial Processes segments, would be pleased to show you the possible applications on site or assist you with your projects, providing competent advice and local support in the practical use and service of the TurBiScat PM 40.azil, would be pleased to show you the possible applications on site or assist you with your projects, providing competent advice and local support in the practical use and service of the TurBiScat PM 40.

request quote
folders
tables

Technical Specifications

Parameter Value
Wavelength turbidity LED 650 nm
Measuring principle turbidity 90°/25° scattered light measurement
Nominal range turbidity 0 … 1,000 EBC (0 … 4,000 NTU)
Resolution turbidity 0.001 EBC of full scale
Reproducibility ±1 % @ 0 .. 2 EBC
±2 % @ 2 .. 100 EBC (90°)
±3 % @ 2 .. 100 EBC (25°)
±10 % @ 100 .. 1,000 EBC
Repeatability 0.001 EBC or ±0.25%
Linearity ±0.5% full scale @ 0 .. 2 EBC
Reaction time less than 2 s (step response – limit value switch)
Wavelength color (optional) LED 430 nm
Nominal range color 0 .. 50 EBC (fix path)
Enclosure Stainless steel 1.4301
Protection class IP 66
Operation voltage 24 VDC ± 10 %
Warm-up time 3 min.
Dimensions Ø 100.5 x 128 mm
Weight Appr. 2.3 kg
Sample temperature -10 .. +100 °C
120 °C max. 2 h
150 °C max. 1 h
Sample pressure – Max. 4 MPa (40 bar) → Sensor head with sapphire windows
– Measurement cell → Please note the specification, standard VARINLINE®-housing 1 MPa (10 bar)
Connections DN 40 … DN 150, 1 1/2″ … 6″ VARINLINE® -housings
Ambient environment -10 .. +50 °C
+55 °C → max. temperature environment +85 °C
+60 °C → max. temperature environment +75 °C
0 .. 100 % relative humidity
Cleaning CIP/SIP compatible up to 120 °C
Material sensor head Stainless steel 1.4404
Windows material Sapphire
Control Units SiDis AD 40

100%

Energy self-sustainable

+4500

Customers around the world

ISO 9001

Certified since 2006

+5000

Parts delivered monthly

Request your quote
Request your quote

    Houve uma falha no envio Envio realizado com sucesso